• Anaesth Intensive Care · Oct 1997

    Randomized Controlled Trial Comparative Study Clinical Trial

    Ventilator-CPAP with the Siemens Servo 900C compared with continuous flow-CPAP in intubated patients: effect on work of breathing.

    • J G Aerts, B van den Berg, and J M Bogaard.
    • Respiratory Intensive Care Unit, University Hospital Dijkzigt, Erasmus University, Rotterdam, The Netherlands.
    • Anaesth Intensive Care. 1997 Oct 1;25(5):487-92.

    AbstractThe effects of continuous positive airway pressure (CPAP) provided by the Siemens Servo 900C ventilator were compared with a continuous flow system (CF-CPAP) in patients weaning from the ventilator. Thirteen patients were studied using both systems at a CPAP level of 0.5 kPa. Additional work of breathing (Wapp) and derived variables were determined in relation to the minute volumes of the patients. The Wapp imposed by the ventilator exceeded the Wapp of CF-CPAP in all patients. The difference in Wapp between ventilator- and CF-CPAP was greater at higher ventilatory needs. The increments in Wapp imposed by the ventilator were positively correlated with the actual end-expiratory pressures (EEP). The EEP increasingly exceeded the preset CPAP level of the ventilator at higher minute volumes. An inspiratory threshold due to a gradient between EEP and preset CPAP greatly increased the Wapp imposed by the ventilator. As this threshold was attributed to the resistance of the PEEP device of the ventilator, it indicates that the additional work related to the expiratory value should be taken into account when the Siemens Servo 900C ventilator is used for weaning purposes.

      Pubmed     Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…