Critical care medicine · Apr 1992
Site of pressure measurement during spontaneous breathing with continuous positive airway pressure: effect on calculating imposed work of breathing.
- M J Banner, R R Kirby, and P B Blanch.
- Department of Anesthesiology, University of Florida College of Medicine, Gainesville.
- Crit. Care Med. 1992 Apr 1;20(4):528-33.
ObjectiveTo describe the importance of measuring pressure at the tracheal end of the endotracheal tube during spontaneous breathing with continuous positive airway pressure in order to correctly assess: a) the changes in airway pressure and b) the work imposed by the breathing apparatus.DesignMultitrial tests under simulated clinical conditions using a mechanical lung model.SettingA research laboratory at a university medical center.InterventionsSpontaneous breathing with continuous positive airway pressure, at peak sinusoidal inspiratory flow-rate demands of 30 and then 60 L/min with sizes 6, 7, 8, and 9 mm internal diameter endotracheal tubes at each flow rate.Measurements And Main ResultsPressure, flow rate, and inhaled and exhaled volumes, during simulated spontaneous ventilation with continuous positive airway pressure were measured. Pressure was measured alternately at the "Y" piece of the breathing tubing of the continuous positive airway pressure system and at the tracheal end of the endotracheal tube to calculate the work imposed by the breathing circuit, endotracheal tube, and the total breathing apparatus. Greater changes in pressure and work were measured at the tracheal end of the endotracheal tube than at the "Y" piece of the breathing tubing for all test conditions. For example, at a peak inspiratory flow-rate demand of 30 L/min when pressures measured at the tracheal end of endotracheal tubes were compared with pressures measured at the "Y"piece, the total work imposed by the breathing apparatus increased by approximately 145% with a 6-mm tube, 95% with a 7-mm tube, 50% with an 8-mm tube, and 40% with a 9-mm tube (p less than .05). Measuring pressure at the "Y" piece of the tubing results in significant underestimations of the changes in pressure and the work imposed, especially when the endotracheal tube has a small internal diameter and/or when the peak inspiratory flow-rate demand is high.ConclusionsThe results indicate that pressure should be measured as close to the patient's airway as possible, i.e., at the tracheal end of the endotracheal tube, rather than using the traditional approach of measuring pressure and assessing work at the inspiratory or expiratory limbs, or "Y" piece of the breathing tubing.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.