Clinical TrialContinuous glucose monitoring system in a rural intensive care unit: a pilot study evaluating accuracy and acceptance.
- Brooke Jacobs, Kim Phan, Leonard Bertheau, Godwin Dogbey, Frank Schwartz, and Jay Shubrook.
- Ohio University College of Osteopathic Medicine, Athens, Ohio 45701 , USA.
- J Diabetes Sci Technol. 2010 May 1;4(3):636-44.
BackgroundGlucose management in an intensive care unit (ICU) is labor-intensive. A continuous glucose monitoring system (CGMS) has the potential to improve efficiency and safety in this setting. The goal of this study was to determine if the Medtronic Guardian REAL-Time CGMS was accurate and tolerated by patients in a rural hospital ICU unit.MethodDifferences between individual finger stick blood glucose (FSBG) and CGMS values were compared to American Diabetes Association (ADA) and International Organization for Standardization (ISO) standards. Continuous glucose monitoring system accuracy was evaluated over four ranges: <75, 75-140, 140-200, and >200 mg/dl. Other accuracy measures [mean absolute deviation (MAD), mean absolute relative difference (MARD), and coefficient of linear regression of CGMS on FSBG] were calculated. Nursing staff and patients were surveyed regarding use of the CGMS in the ICU.ResultsTwenty-nine participants had 320 FSBG and corresponding CGMS readings. Sixty-two percent of participants were admitted with diabetic ketoacidosis (DKA). Two hundred and thirteen (66.6%) were accurate within the ISO standard, whereas only 70 out of 320 (21.9%) were within the 5% ADA standard. The CGMS was most accurate in euglycemia. Technical difficulties, such as adequate time for "wetting" and calibration of electrodes, arose with the sensors. The MAD was 28.3 mg/dl, the MRAD was 17.4%, and the linear regression coefficient of CGMS on FSBG was 0.834 (p < 0.001).ConclusionsThe CGMS is well tolerated by ICU patients but, at present, is not sufficiently accurate to be used for therapeutic decisions in the acute setting, particularly in patients with diabetic ketoacidosis. There is a need to find resolution to the technical issues regarding electrode "wetting" and calibration if CGMS use in the ICU setting is to provide an effective means of diabetes care and management.(c) 2010 Diabetes Technology Society.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.