J Paediatr Child Health · Oct 2011
Comparative StudyProviding PEEP during neonatal resuscitation: which device is best?
- Jennifer A Dawson, Angela Gerber, C Omar F Kamlin, Peter G Davis, and Colin J Morley.
- Neonatal Services, The Royal Women's Hospital, Melbourne, Australia.
- J Paediatr Child Health. 2011 Oct 1;47(10):698-703.
AimThe study aims to compare three commonly used neonatal resuscitation devices, the Laerdal self-inflating bag with a positive end expiratory pressure (PEEP) valve, a T-piece resuscitator (T-piece) and a flow-inflating bag to provide peak inflation pressure (PIP) and PEEP.MethodsParticipants were asked to use each device to give positive pressure ventilation to a modified neonatal mannequin via a face mask to achieve 40-60 inflations per minute, aiming for a PIP/PEEP of 30/5 cm H₂O. A manometer was visible to participants with each device. PIP, PEEP, percentage leak at the face mask and expired tidal volume were measured using a hot-wire anemometer. We analysed 20 inflations from each participant for each device.ResultsFifty participants provided PIP and PEEP with each device. The T-piece was the most accurate and consistent. The flow-inflating bag had the most variation. The leak was lowest with the self-inflating bag and PEEP and highest with the flow-inflating bag, but all had wide variation.ConclusionEach device was able to provide PIP and PEEP when used appropriately. When compared with other resuscitation devices, the T-piece provided the most accurate and consistent PIP and PEEP.© 2011 The Authors. Journal of Paediatrics and Child Health © 2011 Paediatrics and Child Health Division (Royal Australasian College of Physicians).
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.