• Surg Laparosc Endosc Percutan Tech · Apr 2015

    Antenna coupling explains unintended thermal injury caused by common operating room monitoring devices.

    • Nicole T Townsend, Edward L Jones, Alessandro Paniccia, Joel Vandervelde, Jennifer R McHenry, and Thomas N Robinson.
    • *Department of Surgery, University of Colorado School of Medicine, Aurora §Covidien, Boulder, CO †Department of Surgery, The Ohio State University, Columbus, OH ‡University of Kansas School of Medicine, Kansas City, KS.
    • Surg Laparosc Endosc Percutan Tech. 2015 Apr 1;25(2):111-3.

    IntroductionUnintended thermal injury from patient monitoring devices (eg, electrocardiogram pads, neuromonitoring leads) results in third-degree burns. A mechanism for these injuries is not clear. The monopolar "bovie" emits radiofrequency energy that transfers to nearby, nonelectrically active cables or wires without direct contact by capacitive and antenna coupling. The purpose of this study was to determine if, and to what extent, radiofrequency energy couples to common patient monitoring devices.Materials And MethodsIn an ex vivo porcine model, monopolar radiofrequency energy was delivered to a handheld "bovie" pencil. Nonelectrically active neuromonitoring and cardiac-monitoring leads were placed in proximity to the monopolar pencil and its cord. Temperature changes of tissue touched by the monitoring lead were measured using a thermal camera immediately after a 5-second activation. The energy-device cords were then separated by 15 cm, the power was reduced from 30 W coag to 15 W coag and different cord angulation was tested. An advanced bipolar device, a plasma-based device, and an ultrasonic device were also tested at standard settings.ResultsThe neuromonitoring lead increased tissue temperature at the insertion site by 39 ± 13°C (P<0.001) creating visible char at the skin. The electrocardiogram lead raised tissue temperature by 1.3 ± 0.5°C (P<0.001). Decreasing generator power from 30 W to 15 W and separating the bovie cord from the neuromonitoring cord by 15 cm significantly reduced the temperature change (39 ± 13°C vs. 26±5°C; P<0.001 and 39 ± 13°C vs. 10 ± 5°C; P<0.001, respectively). Lastly, monopolar energy increased tissue temperatures significantly more than argon beam energy (34 ± 15°C), advanced bipolar energy (0.2 ± 0.4°C), and ultrasonic energy (0 ± 0.3°C) (all P<0.001).ConclusionsStray energy couples to commonly used patient monitoring devices resulting in potentially significant thermal injury. The handheld bovie cord transfers energy via antenna coupling to neuromonitoring leads that can raise tissue temperatures over 100°F (39°C) using standard settings. The most effective ways to decrease this energy coupling is to reduce generator power, increase the separation between wires, or utilize lower voltage energy devices such as ultrasonic or bipolar energy.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.