    Evaluation of four rapid methods for hemoglobin screening of whole blood donors in mobile collection settings.

    • Antonia Gómez-Simón, Leyre Navarro-Núñez, Elena Pérez-Ceballos, María L Lozano, María J Candela, Almudena Cascales, Constantino Martínez, Javier Corral, Vicente Vicente, and José Rivera.
    • Centro Regional de Hemodonación, C/Ronda de Garay s/n, Unit of Hematology and Clinical Oncology, University of Murcia, 30003 Murcia, Spain.
    • Transfus Apher Sci. 2007 Jun 1;36(3):235-42.

    IntroductionPredonation hemoglobin measurement is a problematic requirement in mobile donation settings, where accurate determination of venous hemoglobin by hematology analyzers is not available.ObjectiveWe have evaluated hemoglobin screening in prospective donors by the semiquantitative copper sulphate test and by capillary blood samples analyzed by three portable photometers, HemoCue, STAT-Site MHgb, and the CompoLab HB system.MethodsCapillary blood samples were obtained from 380 donors and tested by the copper sulphate test and by at least one of the named portable photometers. Predonation venous hemoglobin was also determined in all donors using a Coulter Max-M analyzer.ResultsThe three photometers provided acceptable reproducibility (CV below 5%), and displayed a significant correlation between the capillary blood samples and the venous hemoglobin (R2 0.5-0.8). HemoCue showed the best agreement with venous hemoglobin determination, followed by STAT-Site MHgb, and the CompoLab HB system. The copper sulphate test provided the highest rate of donors acceptance (83%) despite unacceptable hemoglobin levels, and the lowest rate for donor deferral (1%) despite acceptable hemoglobin levels. The percentage of donors correctly categorized for blood donation by the portable hemoglobinometers was 85%, 82%, and 76% for CompoLab HB system, HemoCue and STAT-Site, respectively.ConclusionOur data suggest that hemoglobin determination remains a conflictive issue in donor selection in the mobile setting. Without appropriate performance control, capillary hemoglobin screening by either the copper sulphate method or by the novel portable hemoglobinometers could be inaccurate, thus potentially affecting both donor safety and the blood supply.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

Want more great medical articles?

Keep up to date with a free trial of metajournal, personalized for your practice.
1,694,794 articles already indexed!

We guarantee your privacy. Your email address will not be shared.