• J Clin Monit Comput · Dec 2014

    Dead space reduction by Kolobow's endotracheal tube does not justify the waiving of volume monitoring in small, ventilated lungs.

    • Hans Proquitté, Rena Wendel, Charles C Roehr, Roland R Wauer, and Gerd Schmalisch.
    • Department of Neonatology, Charité University Medical Center, Charitéplatz 1, 10117, Berlin, Germany.
    • J Clin Monit Comput. 2014 Dec 1;28(6):605-11.

    AbstractIn ventilated preterm infants the flow sensor contributes significantly to the total apparatus dead space, which may impair gas exchange. The aim of the study was to quantify to which extent a dead space reduced Kolobow tube (KB) without flow sensor improves the gas exchange compared with a conventional ventilator circuit with flow sensor [Babylog 8000 (BL)]. In a cross-over trial in 14 tracheotomized, surfactant-depleted (saline lavage) and mechanically ventilated newborn piglets (age <12 h; body weight 705-1200 g) BL and KB was applied alternately for 15 min and blood gases were recorded. The inner diameter of the endotracheal tube was 3.6 mm and the apparatus dead space of BL and KB including the endotracheal tube were 3.0 and 1.34 mL. Despite a 50 % apparatus dead space reduction with KB compared to BL statistically significant improvements were only observed for body weights <900 g. In this weight group median paCO2 was decreased by 5 mmHg (p < 0.01), whereas the improvement decreased with decreasing baseline paCO2. Furthermore, median paO2 was increased by 4 mmHg (p < 0.05) and O2 saturation was increased by 2.5 % (p < 0.05). No significant changes were seen in the circulatory parameters. In very small, ventilated lungs the use of KB improved the gas exchange; however, the improvement was moderate and does not justify the waiving of volume monitoring.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.