Clin. Orthop. Relat. Res. · Sep 1994
Comparative StudyBasicervical fractures of the proximal femur. A biomechanical study of 3 internal fixation techniques.
- B Blair, K J Koval, F Kummer, and J D Zuckerman.
- Department of Orthopaedics, Hospital for Joint Diseases, New York City, NY 10003.
- Clin. Orthop. Relat. Res. 1994 Sep 1(306):256-63.
AbstractA biomechanical cadaver study was performed to compare the stability and ultimate strength of 3 standard fixation techniques used for treatment of basicervical hip fractures. Twenty one pairs of mildly osteoporotic femurs were selected, based on a computed tomography bone density reading of 40-50 Hounsfeld units and a Singh index of III. After initial mechanical characterization of intact femurs, basicervical femoral neck fractures were created, reduced, and then instrumented with random assignment to 1 of 3 methods of fixation: (1) 3 parallel 6.5-mm cannulated cancellous screws; (2) a 135 degrees sliding hip screw with a 4 hole side plate; and (3) a 135 degrees sliding hip screw with a 4 hole side plate and a 6.5-mm cannulated cancellous screw placed proximal and parallel to the sliding screw. Nine pairs were tested to failure in axial loading, 6 pairs in lateral bending, and 6 pairs in torsion. The group instrumented with the multiple cancellous screws had a significantly (p < 0.01) lower ultimate axial load to failure than either sliding hip screw group. However, the multiple screws demonstrated significantly (p < 0.01) less fracture displacement. There were no statistically significant differences in lateral bending or torsional testing behavior between the 3 fixation methods. Use of the sliding hip screw is recommended rather than use of multiple cancellous screws for treatment of basicervical femoral neck fractures. Although a superiorly located cancellous screw may provide rotational control during sliding hip screw insertion, it provides no incremental fixation after the sliding hip screw is placed.
Notes
Knowledge, pearl, summary or comment to share?