    Multiple shocks after upgrade of an implantable cardioverter-defibrillator to a cardiac resynchronization therapy-defibrillator device.

    • Gregory M Francisco, Saumya Sharma, Anne H Dougherty, and Bharat K Kantharia.
    • Division of Cardiology, University of Texas, Medical School at Houston, Houston, Texas 77030, USA.
    • Cardiol J. 2009 Jan 1;16(5):473-6.

    AbstractInappropriate shocks from implantable cardioverter-defibrillators (ICDs) may occur for many reasons. Inappropriate shocks are not simply painful inconveniences for patients; they also may result in the need for further operative procedures, and sometimes even death. Herein, we report the case of a patient who after upgrade of an ICD to a cardiac resynchronization therapy-defibrillator device (CRT-D), returned with multiple shocks due to altered sensing and defibrillation polarities that resulted from actual physical reversal of the distal (-) and proximal (+) lead terminals in the header of the device.

      Pubmed     Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…