• Am J Perinatol · Oct 2005

    Comparative Study

    Clinical use of new-generation pulse oximeters in the neonatal intensive care unit.

    • Fegegta A Workie, K Rais-Bahrami, and Billie L Short.
    • Department of Neonatology, Children's National Medical Center, and The George Washington University School of Medicine, Washington, District of Columbia 20010, USA.
    • Am J Perinatol. 2005 Oct 1;22(7):357-60.

    AbstractContinuous monitoring by pulse oximetry is a common practice for preterm and critically ill newborns. A new generation of motion-tolerant pulse oximeters have been designed for improved clinical performance with a substantial reduction in alarm frequency. However, little is known about the differences among these new-generation pulse oximeters in the neonatal intensive care unit (NICU). The purpose of this study is to assess the clinical performance of two new-generation pulse oximeters in the NICU. Two new-generation pulse oximeters were used simultaneously to monitor 36 patients in the NICU. The two devices studied were the Philips FAST and the Masimo SET. Patients were randomly assigned for their digit selection and data were collected only when waveforms were of good quality and/or the pulse oximeter's pulse rate (PR) correlated with the electrocardiogram heart rate (HR). The data for oxygen saturation measurements, number of true and false alarms, and number of dropouts as well as the duration of dropouts for each pulse oximeter were recorded by the primary investigator at 5-minute intervals for a period of 2 hours on each patient. Dropouts are instances when the pulse oximeter alarm sounds due to its inability to identify the arterial pulse and provide an oxygen saturation reading. The mean gestational age for the study group was 32 weeks (rang, 24 to 42 weeks). Repeated-measures analysis of variance indicated no difference between the two devices across all time measurements (p=0.357). In addition, paired t-tests for true alarms and false alarms were not significant, with p-values of 0.151 and 0.869, respectively. There was a difference in the number of data dropouts (p<0.001): the Philips device had a nearly six-fold increase in the number of dropouts (Philips 247 versus Masimo 38). The duration of dropouts was also significant; the Philips device had three times longer duration of dropouts. Physiologic monitoring in the critical care setting requires accurate data measurements. The two new-generation pulse oximeters, the Philips FAST and Masimo SET, are equally sensitive in their ability to identify true and false alarms. Masimo pulse oximeter with its signal extraction technology, however, provides more consistent and accurate reporting of SpO2 values as demonstrated by its markedly decreased incidence of data dropouts. The Masimo unit appears to be more resistant to the effects of motion artifact.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.