- J R Sinclair and I H Wilson.
- University Teaching Hospital, Lusaka, Zambia.
- Anaesthesia. 1988 May 1;43(5):405-8.
AbstractA standard Boyle International anaesthetic machine was modified to allow operation in either a continuous flow or a drawover mode. This was achieved by fitting a valve in the backbar which allows entrainment of air under drawover conditions. The details of the valve and modification are discussed and an evaluation of the machine in a Central African hospital is presented.