J Aerosol Med Pulm Drug Deliv · Aug 2013
In vitro delivery of aerosolized treprostinil via modern mechanical ventilation.
- Rajeev B Patel, Gerald C Smaldone, Ann D Cuccia, and Paul Strachan.
- Stony Brook University Medical Center, Pulmonary, Critical Care and Sleep Medicine, Stony Brook, NY 11794, USA.
- J Aerosol Med Pulm Drug Deliv. 2013 Aug 1;26(4):200-7.
BackgroundThe Tyvaso Inhalation System is a hand-held nebulizer system used to administer treprostinil, an approved therapy for pulmonary arterial hypertension. Our goal was to establish an in vitro method for delivering a standard dose of treprostinil through a ventilator circuit and artificial airway.MethodsAn AeroTech II jet nebulizer (continuous air flow at 10 L/min; Biodex Medical Systems) was placed in the ventilator circuit with a test lung. Two ventilators were tested, the Dräger Evita 2 Dura (Dräger Medical GmbH) and Avea (CareFusion), without humidity. Delivered dose was defined by capturing radiolabeled particles exiting the endotracheal tube with a filter (Pari) and measuring radioactivity. Particle distributions were measured distal to the endotracheal tube by cascade impaction. We hypothesized that drug delivery would be determined by the number of breaths needed, such that the complete time of inspiration totaled 29 sec (e.g., number of breaths needed=29 sec/TI, where TI is the inspiratory time of an average breath read from the ventilator display).ResultsNebulizer output was linear for 6 min, and the standard prescribed target dose of 54 μg (3.1% of full ampule) was delivered in 29 sec. Using our TI algorithm to control delivery, the mean inhaled dose±SD was 72.2±16.5 μg (range 47.2-98.6; n=23). Dräger delivered higher doses than Avea. Effects of mode, breathing pattern, and positive-end expiratory pressures were not significant. The mass median aerodynamic diameter and fine particle fraction were 0.71±0.015 and 0.997±0.0006, respectively.ConclusionsUsing the algorithm, it was possible to deliver aerosolized treprostinil, at controlled doses, via mechanical ventilation over a wide range of controlled breathing patterns. The conditions of nebulization must be precisely followed (one full ampule per treatment, use of the AeroTech II nebulizer, continuous nebulization using an external flow of 10 L/min, bypass of the humidifier or removal of in-line heat and moisture exchanger, and treatment completed in 6 min or less).
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.