Cone beam CT (CBCT) evaluation of inter- and intra-fraction motion for patients undergoing brain radiotherapy immobilized using a commercial thermoplastic mask on a robotic couch.
- A W Lightstone, May Tsao, Parminder S Baran, Gordon Chan, Geordi Pang, Lijun Ma, Fiona Lochray, and Arjun Sahgal.
- Department of Radiation Oncology, Odette Cancer Centre, Sunnybrook Health Sciences Centre, University of Toronto, Toronto, ON, Canada M4N 3M5. Alex.Lightstone@sunnybrook.ca
- Technol. Cancer Res. Treat. 2012 Jun 1;11(3):203-9.
AbstractPatients receiving fractionated intensity-modulated radiation therapy (IMRT) for brain tumors are often immobilized with a thermoplastic mask; however, masks do not perfectly re-orient the patient due to factors including the maximum pressure which can be applied to the face, deformations of the mask assembly, patient compliance, etc. Consequently, ~3-5mm PTV margins (beyond the CTV) are often recommended. We aimed to determine if smaller PTV margins are feasible using mask immobilization coupled with 1) a gantry mounted CBCT image guidance system and 2) position corrections provided by a full six-degree of freedom (6-DOF) robotic couch. A cohort of 34 brain tumor patients was treated with fractionated IMRT. After the mask set-up, an initial CBCT was obtained and registered to the planning CT. The robotic couch corrected the misalignments in all 6-DOF and a pre-treatment verification CBCT was then obtained. The results indicated a repositioning alignment within our threshold of 1.5 mm (3D). Treatment was subsequently delivered. A post-treatment CBCT was obtained to quantify intra-fraction motion. Initial, pre-treatment and post-treatment CBCT image data was analyzed. A total of 505 radiation fractions were delivered to the 34 patients resulting in ~1800 CBCT scans. The initial median 3D (magnitude) set-up positioning error was 2.60 mm. Robotic couch corrections reduced the 3D median error to 0.53 mm prior to treatment. Intra-fraction movement was responsible for increasing the median 3D positioning error to 0.86 mm, with 8% of fractions having a 3D positioning error greater than 2 mm. Clearly CBCT image guidance coupled with a robotic 6-DOF couch dramatically improved the positioning accuracy for patients immobilized in a thermoplastic mask system; however, such intra-fraction motion would be too large for single fraction radiosurgery.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.