- Kozo Kato, Toru Yokoyama, Atsushi Ono, Takuya Numasawa, Kanichiro Wada, Yoshikazu Tonosaki, Yasuhiko Osato, Kazuhiko Sasagawa, and Satoshi Toh.
- Department of Orthopedic Surgery, Hirosaki University Graduate School of Medicine, Japan. kozokato1975@yahoo.co.jp
- J Spinal Disord Tech. 2013 May 1;26(3):E107-11.
Study DesignAn in vitro cadaveric biomechanical study.ObjectiveTo evaluate the anteroposterior (A-P) stability and the flexibility of our novel motion preservation device (MPD) using cadaveric cervical spines.Summary Of Background DataThe MPD intended to restrict the A-P instability of the C1-C2 complex and to preserve the axial rotation, flexion, extension, and lateral bending was designed and produced. The stability and the flexibility of the MPD was evaluated.MethodsTen embalmed cadaveric specimens were loaded with pure A-P translation force and the A-P translational distances were measured. Each specimen was tested for the following 4 models, respectively: Intact (control), the Dens-removed, the MPD instrumented, and a Rod fixation system instrumented. Fifteen specimens were loaded with pure moments (up to 1.5 Nm), and the C1-C2 range of motion (ROM) was measured for flexion, extension, lateral bending, and axial rotation using a stereophotogrammetry motion analysis system.ResultsMean A-P translational distances were 4.26 mm in Intact, 13.1 in the Dens-removed, 5.42 in the MPD, and 2.58 in the Rod fixation. The distance values with the MPD had no significant difference compared with Intact. Mean C1-C2 ROM of Intact, the MPD, and the Rod fixation at 1.5 Nm were: 14.7, 6.96, and 2.11 degrees in flexion, 6.46, 4.72, and 2.84 degrees in extension, 3.29, 4.02, and 1.01 degrees in right lateral bending, 4.92, 4.58, and 1.84 degrees in left lateral bending, 26.4, 15.4, and 1.16 degrees in right axial rotation, and 25.6, 14.3, and 1.21 degrees in left axial rotation, respectively.ConclusionsThe MPD restricted the A-P instability of the C1-C2 complex. In flexion, extension, and axial rotation, the C1-C2 ROM of the MPD was about 50% of the ROM in Intact, whereas equivalent to Intact in lateral bending.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.