Diabetes Technol. Ther. · Oct 2013
Multicenter Study Comparative StudyA new-generation continuous glucose monitoring system: improved accuracy and reliability compared with a previous-generation system.
- Mark Christiansen, Timothy Bailey, Elaine Watkins, David Liljenquist, David Price, Katherine Nakamura, Robert Boock, and Thomas Peyser.
- 1 Diablo Clinical Research , Walnut Creek, California.
- Diabetes Technol. Ther. 2013 Oct 1;15(10):881-8.
BackgroundUse of continuous glucose monitoring (CGM) systems can improve glycemic control, but widespread adoption of CGM utilization has been limited, in part because of real and perceived problems with accuracy and reliability. This study compared accuracy and performance metrics for a new-generation CGM system with those of a previous-generation device.Subjects And MethodsSubjects were enrolled in a 7-day, open-label, multicenter pivotal study. Sensor readings were compared with venous YSI measurements (blood glucose analyzer from YSI Inc., Yellow Springs, OH) every 15 min (±5 min) during in-clinic visits. The aggregate and individual sensor accuracy and reliability of a new CGM system, the Dexcom(®) (San Diego, CA) G4™ PLATINUM (DG4P), were compared with those of the previous CGM system, the Dexcom SEVEN(®) PLUS (DSP).ResultsBoth study design and subject characteristics were similar. The aggregate mean absolute relative difference (MARD) for DG4P was 13% compared with 16% for DSP (P<0.0001), and 82% of DG4P readings were within ± 20 mg/dL (for YSI ≤ 80 mg/dL) or 20% of YSI values (for YSI >80 mg/dL) compared with 76% for DSP (P<0.001). Ninety percent of the DG4P sensors had an individual MARD ≤ 20% compared with only 76% of DSP sensors (P=0.015). Half of DG4P sensors had a MARD less than 12.5% compared with 14% for the DSP sensors (P=0.028). The mean absolute difference for biochemical hypoglycemia (YSI <70 mg/dL) for DG4P was 11 mg/dL compared with 16 mg/dL for DSP (P<0.001).ConclusionsThe performance of DG4P was significantly improved compared with that of DSP, which may increase routine clinical use of CGM and improve patient outcomes.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.