• Med. Sci. Monit. · Jan 2009

    Comparative Study

    Influence of endotracheal tube bore on tidal volume during high frequency oscillatory ventilation: a model lung study.

    • Osamu Hirao, Naoya Iguchi, Akinori Uchiyama, Takashi Mashimo, Masaji Nishimura, and Yuji Fujino.
    • Intensive Care Unit, Osaka University Hospital, Suita, Osaka, Japan.
    • Med. Sci. Monit. 2009 Jan 1;15(1):MT1-4.

    BackgroundIn adults, the influence of endotracheal tube bore on tidal volume (VT) during high-frequency oscillatory ventilation (HFOV) has been little studied.Material/MethodsVia full-length 6, 7, 8, and 9 mm internal diameter (ID) endotracheal tubes (ETTs), a model lung (TTL, Michigan Instruments, MI) was ventilated using a high-frequency oscillator designed for adult patients (R100; Metran, Japan). Oscillation settings were: frequency, 5, 7, and 9 Hz; mean airway pressure, 15 cmH2O; oxygen inspiratory fraction, 0.21. At each setting: oscillator pressure swing was set at maximum; and each test combination was conducted with the model-lung compliance set at 20 and 50 mL/cmH2O. During the trials, a position sensor attached to the model-lung upper surface provided analog electrical signals for lung movement that were converted to VT evaluation data. Pressure and volume signals were digitized using an analog-to-digital converter and recorded in a computer. At each setting, three waveforms were analyzed and the average value used.ResultsAt each setting, measured VT was proportional to ETT internal cross-sectional area (CSA). Regression analysis of VT showed correlation with the CSA, as expressed by formula, y=2.173x where x is CSA and y is VT (R2=0.99, p<0.001) at 5 Hz and compliance of 50 cmH2O.ConclusionsIn an adult-simulating model lung, VT during HFOV varies in proportion to ETT CSA.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

Want more great medical articles?

Keep up to date with a free trial of metajournal, personalized for your practice.
1,694,794 articles already indexed!

We guarantee your privacy. Your email address will not be shared.