    The PediaSat continuous central SvO2 monitoring system does not reliably indicate state or course of central venous oxygenation.

    • Werner Baulig, Dominique Béttex, Christoph Bürki, Achim Schmitz, Nelly Spielmann, Katja Woitzek, and Markus Weiss.
    • Department of Anaesthesia, University Children's Hospital, Switzerland. werner.baulig@usz.ch
    • Eur J Anaesthesiol. 2010 Aug 1;27(8):720-5.

    Background And ObjectiveThe present study compares the accuracy of a new continuous venous oxygenation monitoring system (PediaSat Oximetry Catheter) with laboratory blood oximetry in paediatric surgical patients.MethodsChildren and adolescents undergoing cardiac, orthopaedic or craniofacial surgery with major blood loss were included. A 4.5 Fr two-lumen or 5.5 Fr three-lumen central venous oximetry catheter (SPediaSatcvO2) was inserted preoperatively into the superior vena cava. After in-vivo calibration of the PediaSat system, repeated blood samples were obtained from the distal port of the venous catheter and oximetrically analysed for haemoglobin and central venous oxygen saturation (SCO-OXcvO2). Central venous oxygen saturation values measured by the PediaSat (SPediaSatcvO2) were compared with co-oximetry (SCO-OXcvO2) values from the simultaneously taken blood samples by Bland-Altman and simple regression analyses.ResultsOverall, 142 data pairs from 27 patients, aged from 0.6 to 19.0 years (median 5.3 years) were analysed. SPediaSatcvO2 and SCO-OXcvO2 values ranged from 57 to 98% and from 57.1 to 95.8%, respectively. Correlation between SPediaSatcvO2 and SCO-OXcvO2 was poor with r equal to 0.28 (P < 0.0001). SPediaSatcvO2 overestimated SCO-OXcvO2 (mean bias +2.6%), but limits of agreement (+/-2 SD of bias) were unacceptably high (-14.4/+19.6%). Sensitivity and specificity of SPediaSatcvO2 to indicate a fall or rise of SCO-OXcvO2 between two subsequent measurements were only 0.42 and 0.24, respectively.ConclusionIn paediatric and adolescent patients undergoing major surgery, the PediaSat system did not reliably reflect SCO-OXcvO2 values and cannot replace repeated invasive ScvO2 assessments in the clinically relevant range of ScvO2.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…