    In vitro measurement of flow rate variability in neonatal IV therapy with and without the use of check valves.

    • A C van der Eijk, A J van der Plas, C J N M van der Palen, J Dankelman, and B J Smit.
    • Department of BioMechanical Engineering, Faculty of Mechanical, Maritime & Materials Engineering, Delft University of Technology, Delft, The Netherlands Department of Neonatology, Erasmus MC, Sophia Children's Hospital, University Medical Center Rotterdam, Rotterdam, The Netherlands.
    • J Neonatal Perinatal Med. 2014 Jan 1;7(1):55-64.

    BackgroundIn multi-infusion IV therapy, the actual volume delivered to the neonate can vary over time. To reduce flow rate variability, check valves can be used. A check valve allows flow through the valve in only one direction.ObjectiveTo evaluate flow rate variability in a low flow dual-infusion setup with and without check valves.MethodsThe effect of changing the height of and adding syringes to the IV-administration set was tested with and without check valves in an in vitro dual-infusion setup with in-line flow meters. The pre-programmed flow rates were 2.5 and 0.1 ml/h.ResultsTwenty-four tests of 90 minutes were performed. Time to reach 75% of the pre-programmed 0.1 ml/h flow rate was >20 minutes. The highest total delivered volume during a test was (mean ± SD) 56 ± 8% of the expected delivery for tests without check valves, and diminished to 12 ± 24% of the expected delivery for check valves with a higher opening pressure.ConclusionsThe actual flows and the total delivered volume in low flow dual-infusion setups are less than expected on the pre-programmed flow-rate. These findings emphasize the need for the development of more accurate delivery systems for drugs and fluids in neonatology. Caregivers should be aware of these findings, and optimise the delivery of IV substances by making use of check valves with low opening pressures and by minimising compliance and volume of the IV-administration set. Furthermore, changes in the relative height between pumps and catheter tip should be minimized.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…