- Jen Kok, Jimmy Ng, Stephen C Li, John Giannoutsos, Vineet Nayyar, Jonathan R Iredell, Dominic E Dwyer, and Sharon C-A Chen.
- 1Centre for Infectious Diseases and Microbiology Laboratory Services, Institute of Clinical Pathology and Medical Research, Pathology West, Westmead Hospital, Westmead 2Marie Bashir Institute for Infectious Diseases and Biosecurity, University of Sydney, Westmead Hospital, Westmead 3Centre for Research Excellence in Critical Infections, University of Sydney, Westmead Hospital, Westmead 4Department of Core Pathology and Clinical Chemistry, Institute of Clinical Pathology and Medical Research, Pathology West, Westmead Hospital, Westmead 5Department of Hematology, Institute of Clinical Pathology and Medical Research, Pathology West, Westmead Hospital, Westmead 6Department of Intensive Care Medicine, Westmead Hospital, Westmead, NSW, Australia.
- Pathology. 2015 Aug 1; 47 (5): 405-9.
AbstractData on the performance of point-of-care (POC) or near-patient devices in the management of critically unwell patients are limited, meaning that there are demands for confirming POC test results in the routine clinical laboratory and so potentially leading to delay in treatment provision. We evaluated the performance of the i-STAT CHEM 8+ and CG4+, Hemochron Signature Elite, HemoCue Hb 201+ and WBC Diff Systems on whole blood collected from medical and surgical patients admitted to the intensive care unit at an Australian tertiary care hospital. Measurements obtained for haematology, coagulation, biochemistry and arterial blood gas parameters using POC devices were compared against clinical laboratory analysers (XE-5000, STA-R Evolution, Dimension Vista 1500 and ABL800 FLEX). Bland-Altman and Passing-Bablok regression plots were constructed to assess agreement. Good correlation was defined as a bias of <10% between the POC device and the reference method. Forty arterial blood samples were collected from 28 patients. There was good correlation demonstrated for sodium, potassium, chloride, ionised calcium, glucose, urea, haemoglobin and haematocrit values (i-STAT Chem 8+); pH, pCO(2), bicarbonate and oxygen saturation (i-STAT CG4+); haemoglobin, white cell, neutrophil count and lymphocyte counts (Hemocue); and internationalised normal ratio (INR; Hemochron Signature Elite), but not creatinine, anion gap, pO(2), base excess, lactate, eosinophil count, prothrombin and activated partial thromboplastin time. POC devices were comparable to clinical laboratory analysers in measuring the majority of haematology, biochemistry and coagulation parameters in critically unwell patients, including those with infections. These devices may be deployed at the bedside to allow 'real-time' testing to improve patient care.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.