• Resuscitation · Jun 2014

    Observational Study

    Effect of mattress and bed frame deflection on real chest compression depth measured with two CPR sensors.

    • Marko Sainio, Heidi Hellevuo, Heini Huhtala, Sanna Hoppu, Joar Eilevstjønn, Jyrki Tenhunen, and Klaus T Olkkola.
    • Critical Care Medicine Research Group, Department of Intensive Care Medicine, Tampere University Hospital and University of Tampere, PO Box 2000, FI-33521 Tampere, Finland; Emergency Medical Services, Department of Emergency Medicine, Turku University Hospital, PO Box 52, FI-20521 Turku, Finland. Electronic address: marko.sainio@fimnet.fi.
    • Resuscitation. 2014 Jun 1; 85 (6): 840-3.

    AimImplementation of chest compression (CC) feedback devices with a single force and deflection sensor (FDS) may improve the quality of CPR. However, CC depth may be overestimated if the patient is on a compliant surface. We have measured the true CC depth during in-hospital CPR using two FDSs on different bed and mattress types.MethodsThis prospective observational study was conducted at Tampere University Hospital between August 2011 and September 2012. During in-hospital CPR one FDS was placed between the rescuer's hand and the patient's chest, with the second attached to the backboard between the patient's back and the mattress. The real CC depth was calculated as the difference between the total depth from upper FDS to lower FDS.ResultsTen cardiac arrests on three different bed and mattress types yielded 10,868 CCs for data analyses. The mean (SD) mattress/bed frame effect was 12.8 (4) mm on a standard hospital bed with a gel mattress, 12.4 (4) mm on an emergency room stretcher with a thin gel mattress and 14.1 (3) mm on an ICU bed with an emptied air mattress. The proportion of CCs with an adequate depth (≥50 mm) decreased on all mattress types after compensating for the mattress/bed frame effect from 94 to 64%, 98 to 76% and 91 to 17%, in standard hospital bed, emergency room stretcher and ICU bed, respectively (p<0.001).ConclusionThe use of FDS without real-time correction for deflection may result in CC depth not reaching the recommended depth of 50 mm.Copyright © 2014 Elsevier Ireland Ltd. All rights reserved.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…