Int J Clin Monit Comput · Aug 1994
Comparative StudyPresentation and evaluation of a new optical sensor for respiratory rate monitoring.
- M Vegfors, L G Lindberg, H Pettersson, and P A Oberg.
- Department of Anaesthesiology, Linköping University, Sweden.
- Int J Clin Monit Comput. 1994 Aug 1;11(3):151-6.
AbstractA new optical sensor for respiratory rate monitoring was simultaneously compared with an acoustic sensor and a transthoracic impedance plethysmograph during normoventilation in the respiratory rate range of 9-17 breaths per minute. The response characteristics of the optical sensor were then measured during simulation of central apnoea and tachypnoea. Visual observation was chosen as the reference method for monitoring the respiratory rate. The measurements were performed in ten healthy volunteers and the respiratory signals recorded on an analogue tape and strip-chart recorder and analysed off-line. The response characteristics of the fibre optic sensor corresponded well with those of the acoustic sensor and impedance plethysmograph. All three methods responded rapidly to an apnoeic event.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.