• Clin Chem Lab Med · Jan 2006

    Evaluation and quality assessment of glucose concentration measurement in blood by point-of-care testing devices.

    • Michael Wehmeier, Burkhard T Arndt, Gerhard Schumann, and Wolf R Külpmann.
    • Medical School Hannover, Department of Clinical Chemistry, Hannover, Germany. wehmeier.michael@mh-hannover.de
    • Clin Chem Lab Med. 2006 Jan 1;44(7):888-93.

    BackgroundIn this study the analytical performance of eight glucose point-of-care testing (POCT) devices was evaluated. For this purpose, POCT measurement of glucose in heparinized blood collected from patients was paralleled by determination of the glucose concentration in the respective plasma by an analyzer (Hitachi 917) in the central laboratory, providing traceable results.MethodsTrueness of POCT measurements was studied by comparing the plasma POCT values (mean of five measurements) with the results from the traceable measurement procedure (TMP).ResultsThe percentage of POCT results within +/-6% of the TMP mean value ranged from 24% to 50%, depending on the POCT device. Within the reference interval of plasma glucose (4.4-6.0 mmol/L), up to 67% of the POCT values were lower than 4.4 mmol/L, leading to a false diagnosis of hypoglycemia. In the hypoglycemic range (<4.4 mmol/L) up to 29% of the POCT analyses were false normoglycemic.ConclusionsIn conclusion, this study shows an insufficient trueness of glucose measurements by POCT devices in the normo- and hypoglycemic range. To improve quality assessment, sample splitting and simultaneous measurement of blood glucose concentration every 4 weeks by POCT devices and of plasma glucose concentration by a reliable TMP is recommended.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…