• Crit Care Resusc · Jun 2007

    Effect of changes in syringe driver height on flow: a small quantitative study.

    • Andrew I Donald, Meher P Chinthamuneedi, and Douglas Spearritt.
    • Intensive Care Unit, St Vincent's Hospital, Toowoomba, QLD, Australia.
    • Crit Care Resusc. 2007 Jun 1;9(2):143-7.

    ObjectiveTo quantify flow irregularities in drug delivery caused by vertical displacement of syringe pumps.MethodsA bench experiment was performed to quantify the effect of height on pumps used in our intensive care and theatre settings. A standard syringe pump and line set loaded with a dye solution was run through a graduated length of tubing, and the effect of changing pump height quantified by measuring progress down the tubing over time.ResultsA 30 cm elevation produced significant drug delivery boluses - up to seven times the programmed rate at 2mL/h. Delivery rate increased in inverse proportion to the programmed rate, as did the time taken to return to the programmed rate. Lowering the pump 30 cm resulted in no-flow times of up to 180 seconds at a flow rate of 2mL/h - again inversely proportional to programmed rate.ConclusionsVertical displacement of a common syringe pump by 30 cm produced significant bolus and cessation phenomena. These findings confirm the observations of previous authors and also demonstrate significant flow irregularities with smaller vertical displacements than previously tested. Further testing with other brands of pumps is required before a solution to this clinically important problem may be approached.

      Pubmed     Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.