• J Clin Monit Comput · Feb 2015

    Observational Study

    The relative trending accuracy of noninvasive continuous hemoglobin monitoring during hemodialysis in critically ill patients.

    • Hiroshi Yamada, Minako Saeki, Junko Ito, Kazuhiro Kawada, Aya Higurashi, Hiromi Funakoshi, and Kohji Takeda.
    • Critical Care Division, Fujisawa City Hospital, 2-6-1 Fujisawa, Fujisawa, Kanagawa, 251-8550, Japan, hiroymd52@gmail.com.
    • J Clin Monit Comput. 2015 Feb 1;29(1):107-12.

    AbstractThe pulse CO-Oximeter (Radical-7; Masimo Corp., Irvine, CA) is a multi-wavelength spectrophotometric method for noninvasive continuous monitoring of hemoglobin (SpHb). Because evaluating the relative change in blood volume (ΔBV) is crucial to avoid hypovolemia and hypotension during hemodialysis, it would be of great clinical benefit if ΔBV could be estimated by measurement of SpHb during hemodialysis. The capability of the pulse CO-Oximeter to monitor ΔBV depends on the relative trending accuracy of SpHb. The purpose of the current study was to evaluate the relative trending accuracy of SpHb by the pulse CO-Oximeter using Crit-Line as a reference device. In 12 patients who received hemodialysis (total 22 sessions) in the intensive care unit, ΔBV was determined from SpHb. Relative changes in blood volume determined from SpHb were calculated according to the equation: ΔBV(SpHb)=[starting SpHb]/[current SpHb] - 1. The absolute values of SpHb and hematocrit measured by Crit-Line (CL-Hct) showed poor correlation. On the contrary, linear regression analysis showed good correlation between ΔBV(SpHb) and the relative change in blood volume measured by Crit-Line [ΔBV(CL-Hct)] (r=0.83; P≤0.001). Bland-Altman analysis also revealed good agreement between ΔBV(SpHb) and ΔBV(CL-Hct) (bias, -0.77%; precision, 3.41%). Polar plot analysis revealed good relative trending accuracy of SpHb with an angular bias of 4.1° and radial limits of agreement of 24.4° (upper) and -16.2° (lower). The results of the current study indicate that SpHb measurement with the pulse CO-Oximeter has good relative trending accuracy.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

Want more great medical articles?

Keep up to date with a free trial of metajournal, personalized for your practice.
1,694,794 articles already indexed!

We guarantee your privacy. Your email address will not be shared.