• Pacing Clin Electrophysiol · Oct 1998

    Case Reports Comparative Study

    Electromagnetic interference from a muscle stimulation device causing discharge of an implantable cardioverter defibrillator: epicardial bipolar and endocardial bipolar sensing circuits are compared.

    • T V Glotzer, M Gordon, M Sparta, G Radoslovich, and J Zimmerman.
    • Department of Medicine, Hackensack University Medical Center, New Jersey 07601, USA.
    • Pacing Clin Electrophysiol. 1998 Oct 1;21(10):1996-8.

    AbstractThis case report is about two patients with two different types of ICDs who underwent electrical muscle stimulation (EMS) therapy. In one patient with an ICD that has epicardial screw-in bipolar sensing leads, electromagnetic interference (EMI) from the EMS device caused the delivery of an inappropriate ICD discharge. In a second patient with an ICD with endocardial true bipolar sensing, there was no evidence of EMI during the EMS therapy despite all of our attempts to reproduce it. The sensing circuits in the two different ICDs are compared.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…