• Critical care medicine · Feb 1987

    Evaluation of an emergency cricothyrotomy instrument.

    • D G Bjoraker, N B Kumar, and A C Brown.
    • Crit. Care Med. 1987 Feb 1;15(2):157-60.

    AbstractAn emergency cricothyrotomy device was placed in 11 anesthetized dogs in order to assess airway damage and problems in placement, ventilation, and design. Posterior airway perforation without esophageal damage occurred in three animals. Submucosal hematomas or cricoid cartilage injury occurred in seven animals. As placement by an untrained operator was surprisingly difficult, prior operator training with this device is recommended. When properly installed, an excellent emergency airway was achieved. Some minor design shortcomings are noted and suggestions for improvement are made.

      Pubmed     Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.