• Blood pressure monitoring · Feb 1999

    Evaluation of the Schiller BR-102 ambulatory blood pressure system according to the protocols of the British Hypertension Society and the Association for the Advancement of Medical Instrumentation.

    • E O'Brien, F Mee, and N Atkins.
    • The Blood Pressure Unit, Beaumont Hospital, Dublin 9, Ireland.
    • Blood Press Monit. 1999 Feb 1;4(1):35-43.

    ObjectiveTo evaluate the Schiller BR-102 monitor for ambulatory blood pressure measurement according to the protocols of the British Hypertension Society (BHS) and the Association for the Advancement of Medical Instrumentation (AAMI).DesignThe BHS protocol is divided into two parts. Part I, which is the part applicable to this study, comprises the main validation procedure and has five phases: before-use device calibration; in-use (field) phase; after-use device calibration; static device validation; and report of evaluation.MethodThree Schiller BR-102 recorders passed the before-use device calibration test, after which they entered the in-use (field) assessment phase during which the three recorders were each worn by 10 subjects for 24 h, after which calibration was again assessed. Because there was no difference in results of calibration testing among the three devices, one was selected randomly and the main validation test was carried out on 85 subjects with a wide range of blood pressures both for the auscultatory mode and for the oscillometric mode using the Sphygmocorder. The results were analysed according to the BHS grading system from A to D. The data were also analysed according to the standard of the Association for the Advancement of Medical Instrumentation (AAMI), which stipulates that the mean difference between the test device and the standard shall be ResultsThe Schiller BR-102 achieved a BHS grade B rating for systolic and diastolic blood pressures in the auscultatory mode and satisfied the criteria for accuracy of the AAMI protocol for systolic and diastolic blood pressures. In the oscillometric mode, the Schiller BR-102 achieved grade D for systolic blood pressure and grade B for diastolic blood pressure according to the BHS protocol and satisfied the AAMI criteria for diastolic but not systolic blood pressure. Applying the BHS and AAMI criteria to tertiles of blood pressure (low-pressure range < 130/80 mmHg, medium-pressure range 130-160/80-100 mmHg, high-pressure range > 160/100 mmHg) the Schiller BR-102 was less accurate in the high pressure range for diastolic blood pressure but more accurate for systolic blood pressure, achieving A/C grading, while satisfying the AAMI criteria both for systolic and for diastolic blood pressure in the auscultatory mode. In the oscillometric mode the device performed less accurately in the high-pressure range, achieving grade D/C, while failing to satisfy the AAMI criteria both for systolic and for diastolic blood pressure. The means+/-SD of the first mercury sphygmomanometer measurements were 143+/-32 mmHg for systolic blood pressure and 88+/-21 mmHg for diastolic blood pressure. Acceptability to subjects was good and the manufacturer's manual was satisfactory.ConclusionOn the basis of these results, the Schiller BR-102 can be recommended for ambulatory blood pressure measurement in clinical practice using the auscultatory mode, but the oscillometric mode, which operates only if the device fails in the auscultatory mode, does not provide accurate measurements.

      Pubmed     Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…