• Anaesthesia · Jun 2013

    Pressure changes during tracheal suctioning--a laboratory study.

    • S G C Palazzo and B Soni.
    • Chelsea and Westminster Hospital, London, UK. sgcpalazzo@gmail.com
    • Anaesthesia. 2013 Jun 1;68(6):576-84.

    AbstractSuctioning is essential in managing tracheal tubes, but also has drawbacks. Using a bench model, we demonstrated the extent and time course of pressure changes during suctioning, examined their relationship with tracheal tube and catheter diameters and assessed the effects of artificial 'sputum' and of compensatory gas flow in the system. We suctioned at -20 kPa (-150 mmHg) and -80 kPa (-600 mmHg) using three different sized catheters and a 5.9-mm diameter bronchoscope through tracheal tubes ranging from 6.5 mm to 9.0 mm in diameter. Pressure changes ranged from -0.1 kPa (-0.8 mmHg) to -20.4 kPa (-153.0 mmHg). We demonstrated more negative pressures with decreasing tracheal tube diameter (p = 0.024) and increasing catheter diameter (p = 0.038). Addition of artificial 'sputum' led to more negative, but unpredictable, pressures than those seen with clean tubes (p = 0.012). Bronchoscopic suctioning produced pressure changes even greater than the largest suction catheter (p = 0.0039). Using a closed system with continuous positive airway pressure and 155 l.min(-1) compensatory gas flow attenuated the pressure changes generated both with a 4.0-mm catheter (p = 0.0005) and on bronchoscopic suctioning (p = 0.0078). The time taken to reach 50% of minimum pressure was always less than 1 s. The use of high compensatory flows during suctioning merits clinical evaluation.Anaesthesia © 2013 The Association of Anaesthetists of Great Britain and Ireland.

      Pubmed     Free full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…