• Resuscitation · Nov 2001

    Comparative Study

    Pediatric transthoracic defibrillation: biphasic versus monophasic waveforms in an experimental model.

    • C B Clark, Y Zhang, L R Davies, G Karlsson, and R E Kerber.
    • The Cardiovascular Center, The University of Iowa, Iowa City, IA 52242, USA.
    • Resuscitation. 2001 Nov 1;51(2):159-63.

    ObjectivesThe purpose of this study was to determine and compare the efficacy of biphasic and monophasic waveforms in a porcine model of pediatric defibrillation.BackgroundThe efficacy and safety of biphasic waveforms in children has not been established.MethodsWe initially studied 27 piglets: 12 weighed 3-6 kg ('infants'), and 15 weighed 7-12 kg ('children'). Ventricular fibrillation (VF) was induced by rapid right ventricular pacing and maintained for 15 s. Transthoracic shocks of 7-100 J energy were given using monophasic (5 ms truncated exponential) and biphasic (5 ms positive, 5 ms negative pulse, truncated exponential) waveforms. A second study of four 'infant' and four 'child' piglets utilized the same protocol but with a 10 ms instead of 5 ms monophasic truncated exponential shock waveform compared with the 10 ms biphasic waveform.ResultsFor both biphasic and monophasic waveforms, shock success rate (termination of VF) rose steadily as energy was increased. In the first study in the 'infant' 3-6 kg group, the 10 ms biphasic waveforms were superior to 5 ms monophasic waveforms at 10, 20, and 30 J energies, and in the 'child' 7-12 kg group at 20 and 30 J energies (P<0.05). High success rates (>80%) were achieved by 20 J (4 J/kg) biphasic waveform shocks in the 'infant' piglets and 30 J (3 J/kg) biphasic waveform shocks in the 'child' piglets. In the second study using a 10 ms monophasic waveform, we found similar results. Pulseless electrical activity occurred in two animals following biphasic shocks and in two animals following monophasic shocks.ConclusionsBiphasic waveforms proved superior to monophasic waveforms in both infant and child models. High success rates were achieved with low-energy biphasic shocks. Biphasic waveform defibrillation is a promising advance in pediatric resuscitation.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…