-
Anesthesia and analgesia · Jun 2014
Automated checkout routines in anesthesia workstations vary in detection and management of breathing circuit obstruction.
- Michael P Dosch.
- From the Department of Nurse Anesthesia, College of Health Professions, University of Detroit Mercy, Detroit, Michigan.
- Anesth. Analg.. 2014 Jun 1;118(6):1254-7.
AbstractWhile rare, anesthesia breathing system obstruction can have devastating consequences. We created simulated occlusions of the expiratory and inspiratory limb of the circle breathing system in 3 current anesthesia workstations; Aisys, ADU (both by GE Healthcare, Madison WI), and Apollo (Draeger Medical, Telford, PA). The automated electronic checkout specific to each machine was then performed. The Aisys allowed users to accept both faults and initiate simulated patient care; the ADU and Apollo did not. Users must be aware of how to test for breathing circuit obstruction, and whether their own equipment does so adequately in the automated checkout.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.