    Accuracy of a new real-time continuous glucose monitoring algorithm.

    • D Barry Keenan, Raymond Cartaya, and John J Mastrototaro.
    • Medtronic MiniMed, Northridge, California 91325, USA. barry.keenan@medtronic.com
    • J Diabetes Sci Technol. 2010 Jan 1;4(1):111-8.

    BackgroundThrough minimally invasive sensor-based continuous glucose monitoring (CGM), individuals can manage their blood glucose (BG) levels more aggressively, thereby improving their hemoglobin A1c level, while reducing the risk of hypoglycemia. Tighter glycemic control through CGM, however, requires an accurate glucose sensor and calibration algorithm with increased performance at lower BG levels.MethodsSensor and BG measurements for 72 adult and adolescent subjects were obtained during the course of a 26-week multicenter study evaluating the efficacy of the Paradigm REAL-Time (PRT) sensor-augmented pump system (Medtronic Diabetes, Northridge, CA) in an outpatient setting. Subjects in the study arm performed at least four daily finger stick measurements. A retrospective analysis of the data set was performed to evaluate a new calibration algorithm utilized in the Paradigm Veo insulin pump (Medtronic Diabetes) and to compare these results to performance metrics calculated for the PRT.ResultsA total of N = 7193 PRT sensor downloads for 3 days of use, as well as 90,472 temporally and nonuniformly paired data points (sensor and meter values), were evaluated, with 5841 hypoglycemic and 15,851 hyperglycemic events detected through finger stick measurements. The Veo calibration algorithm decreased the overall mean absolute relative difference by greater than 0.25 to 15.89%, with hypoglycemia sensitivity increased from 54.9% in the PRT to 82.3% in the Veo (90.5% with predictive alerts); however, hyperglycemia sensitivity was decreased only marginally from 86% in the PRT to 81.7% in the Veo.ConclusionsThe Veo calibration algorithm, with sensor error reduced significantly in the 40- to 120-mg/dl range, improves hypoglycemia detection, while retaining accuracy at high glucose levels.2010 Diabetes Technology Society.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…