• Anaesth Intensive Care · Feb 2003

    Consequences of syringe size sensor malfunction in a modern infusion pump.

    • J L Derrick, A M H Ho, and A M W Cho.
    • Department of Anaesthesia and Intensive Care, Prince of Wales Hospital, Faculty of Medicine, The Chinese University of Hong Kong, Shatin, Hong Kong.
    • Anaesth Intensive Care. 2003 Feb 1;31(1):75-9.

    AbstractPrompted by an actual case of potentially life-threatening infusion pump malfunction, we investigated the effects of wire breakage(s) within the syringe size sensor circuit in a Graseby 3400 infusion pump. The circuit wires within the sensor were systematically broken. The syringe sizes recognised by the sabotaged circuit and the actual sizes of syringes inserted into the pump were compared. Thirty-eight per cent of the possible wire breakages resulted in a smaller syringe size being recognized, causing the infusion rate to be too fast, and 38% of the possiblewire breakage resulted in a larger syringe size being recognized, causing the infusion rate to be too slow. The volume delivered for each different size of Terumo syringe as a function of distance travelled by the plunger was measured. The errors ranged from 0.4 to 2.6 times that of the expected rate. Only 1.3% of the possible wire breakage(s) were recognised as errors by the pump. The infusion rates were not affected in 22.5% of the cases. Wire breakage within the syringe size sensor in infusion pumps is yet another potential source of infusion error, with important safety implications.

      Pubmed     Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.