Critical care medicine · Jun 2016
Observational StudyEvaluation of Noninvasive Hemoglobin Monitoring in Surgical Critical Care Patients.
- Tim Xu, Ting Yang, Joseph B Kim, Mark C Romig, Adam Sapirstein, and Bradford D Winters.
- 1Department of Anesthesiology and Critical Care Medicine, Johns Hopkins University School of Medicine, Johns Hopkins University, Baltimore, MD. 2Armstrong Institute for Patient Safety and Quality, Johns Hopkins University School of Medicine, Johns Hopkins University, Baltimore, MD.
- Crit. Care Med. 2016 Jun 1; 44 (6): e344-52.
ObjectiveTo assess the clinical utility of noninvasive hemoglobin monitoring based on pulse cooximetry in the ICU setting.Design And SettingA total of 358 surgical patients from a large urban, academic hospital had the noninvasive hemoglobin monitoring pulse cooximeter placed at admission to the ICU. Core and stat laboratory hemoglobin measurements were taken at the discretion of the clinicians, who were blinded to noninvasive hemoglobin monitoring values.Measurement And Main ResultsThere was a poor correlation between the 2,465 time-matched noninvasive hemoglobin monitoring and laboratory hemoglobin measurements (r = 0.29). Bland-Altman analysis showed a positive bias of 1.0 g/dL and limits of agreement of -2.5 to 4.6 g/dL. Accuracy was best at laboratory values of 10.5-14.5 g/dL and least at laboratory values of 6.5-8 g/dL. At hemoglobin values that would ordinarily identify a patient as requiring a transfusion (< 8 g/dL), noninvasive hemoglobin monitoring consistently overestimated the patient's true hemoglobin. When sequential laboratory values declined below 8 g/dL (n = 102) and 7 g/dL (n = 13), the sensitivity and specificity of noninvasive hemoglobin monitoring at identifying these events were 27% and 7%, respectively. At a threshold of 8 g/dL, continuous noninvasive hemoglobin monitoring values reached the threshold before the labs in 45 of 102 instances (44%) and at 7 g/dL, noninvasive hemoglobin monitoring did so in three of 13 instances (23%). Noninvasive hemoglobin monitoring minus laboratory hemoglobin differences showed an intraclass correlation coefficient of 0.47 within individual patients. Longer length of stay and higher All Patient Refined Diagnostic-Related Groups severity of illness were associated with poor noninvasive hemoglobin monitoring accuracy.ConclusionsAlthough noninvasive hemoglobin monitoring technology holds promise, it is not yet an acceptable substitute for laboratory hemoglobin measurements. Noninvasive hemoglobin monitoring performs most poorly in the lower hemoglobin ranges that include commonly used transfusion trigger thresholds and is not consistent within individual patients. Further refinement of the signal acquisition and analysis algorithms and clinical reevaluation are needed.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.