Anesthesia and analgesia · Jan 1994
Comparative StudySyringe pumps for infusion of vasoactive drugs: mechanical idiosyncrasies and recommended operating procedures.
- G A Rooke and T A Bowdle.
- Department of Anesthesiology, University of Washington School of Medicine, Seattle.
- Anesth. Analg. 1994 Jan 1;78(1):150-6.
AbstractSyringe pumps for vasoactive infusions have the advantages of small size and weight, portability, and low cost of the disposable components. However, limited syringe capacity necessitates the use of high drug concentrations, and the accidental delivery of even a small volume of infusate could seriously alter the patient's hemodynamics. To determine the circumstances under which drug delivery might be delayed, or inadvertent boluses could be delivered into the manifold, two brands of commercially available clinical syringe pumps were connected to a stopcock manifold via small-bore tubing and a series of tests were performed. When the syringe pumps were operated at 3 mL/h against a closed stopcock, the pumps' occlusion alarms did not sound for 18-22 min, and when the stopcock subsequently was opened, 0.6-0.9 mL of infusate was delivered as a bolus. Elevating the syringe pump by 120 cm resulted in the delivery of up to 0.5 mL of infusate with the pump turned off. When a syringe pump operating at 6 mL/h was turned off, typically an additional 0.05 mL was delivered during the ensuing 2-3 min. Depending upon the method used to flush the tubing prior to use, delays in drug delivery of 2-3 min occurred at an infusion rate of 3 mL/h. These observations emphasize the need for careful equipment setup and proper use of the manifold stopcocks to avoid unintended drug administration or delay in drug administration.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.