    The pressure exerted on the tracheal wall by two endotracheal tube cuffs: A prospective observational bench-top, clinical and radiological study.

    • Alex Doyle, Ramai Santhirapala, Martin Crowe, Mark Blunt, and Peter Young.
    • Critical Care Trainee, Department of Anaesthesia and Critical Care, Queen Elizabeth Hospital, Kings Lynn, Norfolk, UK, PE30 4ET. ajdoyle@gmail.com.
    • BMC Anesthesiol. 2010 Jan 1;10:21.

    BackgroundThe Lotrach endotracheal tube has a unique low-volume, low-pressure (LVLP) cuff, which has been designed to prevent pressure injury to the tracheal wall. We aimed to estimate the pressure exerted on the tracheal wall by the LVLP cuff and a conventional cuff in a bench-top, clinical and radiological study.MethodIn the bench-top study, a model trachea was intubated with the LVLP cuff and the conventional cuff. The cuff pressure was controlled using a constant pressure device. We assessed the pressure exerted on the tracheal wall by measuring the ability of the cuffs to support a column of water using a standard protocol. In the clinical study, we tested the ability of both cuffs to prevent air leak during a staged recruitment manoeuvre. In the radiological study, we recorded the degree of anatomical distortion of the trachea from both cuffs in the antero-posterior (AP) and transverse tracheal diameters. We performed statistical analysis using non-inferiority tests.ResultsIn the bench-top study, the LVLP cuff achieved a plateau at a mean height of 25.2 cmH2O (SD 0.34). In contrast, the conventional cuff failed to maintain any water above the cuff and a plateau could not be measured. In the clinical study, the mean pressure at which air leak occurred was 30.0 +/- 0.8 cmH2O (SD 3.8) using the LVLP cuff and 32.4 +/- 0.7 cmH2O (SD 3.0) using the conventional cuff. In the radiological study, the mean degree of anatomical distortion of the trachea in AP and transverse tracheal diameter was 2.9 +/- 2.2 mm (SD 2.1) and 1.8 +/- 1.4 mm (SD 1.4) using the LVLP cuff and 4.4 +/- 1.3 mm (SD 1.4) and 2.6 +/- 1.5 mm (SD 1.6) using the conventional cuff.ConclusionsThe bench-top and clinical studies both demonstrated that the LVLP cuff exerted approximately 30 cmH2O of pressure on the tracheal wall. These results are supported by our radiological study. We conclude that the LVLP cuff exerts an acceptable amount of pressure on the tracheal wall when it is operated at the recommended intracuff pressure.

      Pubmed     Free full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.