J Clin Monit Comput · Apr 2015
Observational StudyThe evaluation of a non-invasive respiratory volume monitor in surgical patients undergoing elective surgery with general anesthesia.
- Christopher J Voscopoulos, C Marshall MacNabb, Jordan Brayanov, Lizeng Qin, Jenny Freeman, Gary John Mullen, Diane Ladd, and Edward George.
- Department of Anesthesiology, Pain and Perioperative Medicine, Brigham and Woman's Hospital, Harvard Medical School, Boston, MA, USA, cjvoscopoulos@yahoo.com.
- J Clin Monit Comput. 2015 Apr 1; 29 (2): 223-30.
AbstractContinuous respiratory assessment is especially important during post-operative care following extubation. Respiratory depression and subsequent adverse outcomes can arise due to opioid administration and/or residual anesthetics. A non-invasive respiratory volume monitor (RVM) has been developed that provides continuous, real-time, measurements of minute ventilation (MV), tidal volume (TV), and respiratory rate (RR) via a standardized set of thoracic electrodes. Previous work demonstrated accuracy of the RVM versus standard spirometry and its utility in demonstrating response to opioids in postoperative patients. This study evaluated the correlation between RVM measurements of MV, TV and RR to ventilator measurements during general anesthesia (GA). Continuous digital RVM and ventilator traces, as well as RVM measurements of MV, TV and RR, were analyzed from ten patients (mean 62.6±7.4 years; body mass index 28.6±5.2 kg/m2) undergoing surgery with GA. RVM data were compared to ventilator data and bias, precision and accuracy were calculated. The average MV difference between the RVM and ventilator was -0.10 L/min (bias: -1.3%, precision: 6.6%, accuracy: 9.0%. The average TV difference was 40 mL (bias: 0.4%, precision: 7.3%, accuracy: 9.1%). The average RR difference was -0.22 breaths/minute (bias: -1.8%, precision: 3.7% accuracy: 4.1%). Correlations between the RVM traces and the ventilator were compared at various points with correlations>0.90 throughout. Pairing the close correlation to ventilator measurements in intubated patients demonstrated by this study with previously described accuracy compared to spirometry in non-intubated patients, the RVM can be considered to have the capability to provide continuity of ventilation monitoring post-extubation This supports the use of real-time continuous RVM measurements to drive post-operative and post-extubation protocols, initiate therapeutic interventions and improve patient safety.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.