• Anesthesia and analgesia · Sep 2014

    Comparative Study

    An Evaluation of a Zero-Heat-Flux Cutaneous Thermometer in Cardiac Surgical Patients.

    • Yashar Eshraghi, Vivian Nasr, Ivan Parra-Sanchez, Albert Van Duren, Mark Botham, Thomas Santoscoy, and Daniel I Sessler.
    • From the Departments of *Outcomes Research, †General Anesthesiology, Cleveland Clinic, Cleveland, Ohio; §Infection Prevention Division, The 3M Company, St. Paul, Minnesota.
    • Anesth. Analg.. 2014 Sep 1;119(3):543-9.

    BackgroundAlthough core temperature can be measured invasively, there are currently no widely available, reliable, noninvasive thermometers for its measurement. We thus compared a prototype zero-heat-flux thermometer with simultaneous measurements from a pulmonary artery catheter. Specifically, we tested the hypothesis that zero-heat-flux temperatures are sufficiently accurate for routine clinical use.MethodsCore temperature was measured from the thermistor of a standard pulmonary artery catheter and with a prototype zero-heat-flux deep-tissue thermometer in 105 patients having nonemergent cardiac surgery. Zero-heat-flux probes were positioned on the lateral forehead and lateral neck. Skin surface temperature probes were attached to the forehead just adjacent to the zero-heat-flux probe. Temperatures were recorded at 1-minute intervals, excluding the period of cardiopulmonary bypass, and for the first 4 postoperative hours. Zero-heat-flux and pulmonary artery temperatures were compared with bias analysis; differences exceeding 0.5°C were considered to be potentially clinically important.ResultsThe mean duration in the operating room was 279 ± 75 minutes, and the mean cross-clamp time was 118 ± 50 minutes. All subjects were monitored for an additional 4 hours in the intensive care unit. The average overall difference between forehead zero-heat-flux and pulmonary artery temperatures (i.e., forehead minus pulmonary artery) was -0.23°C (95% limits of agreement of ±0.82); 78% of the differences were ≤0.5°C. The average intraoperative temperature difference was -0.08°C (95% limits of agreement of ±0.88); 84% of the differences were ≤0.5°C. The average postoperative difference was -0.32°C (95% limits of agreement of ±0.75); 84% of the differences were ≤0.5°C. Bias and precision values for neck site were similar to the forehead values. Uncorrected forehead skin temperature showed an increasing negative bias as core temperature decreased.ConclusionsCore temperature can be noninvasively measured using the zero-heat-flux method. Bias was small, but precision was slightly worse than our designated 0.5°C limits compared with measurements from a pulmonary artery catheter.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.