- Yan Lu, John Bogdanske, Monica Lopez, Brian J Cole, and Mark D Markel.
- Comparative Orthopaedic Research Laboratory, Department of Medical Sciences, School of Veterinary Medicine, University of Wisconsin-Madison, Madison, Wisconsin 53706-11102, USA.
- Arthroscopy. 2005 May 1;21(5):592-6.
PurposeTo determine joint fluid temperatures at different time intervals during treatment with radiofrequency energy (RFE) applied in intermittent and continuous treatment manners under flow or no-flow conditions using a simulated shoulder joint model.Type Of StudyIn vitro measurement of simulated joint fluid temperature during RFE treatment.MethodsA custom-built jig with a chamber (volume size, 25 mL) was used to mimic the adult human shoulder. Three RFE systems: Vulcan EAS plus TAC-S probe (Smith & Nephew Endoscopy, Andover, MA); VAPR II plus End-Effect Electrode (Mitek, Westwood, MA); and ArthroCare 2000 plus TurboVac 90 degrees probe (ArthroCare, Sunnyvale, CA) were tested in the chamber with saline solution initially set at 23 degrees C. Each RFE probe was applied in a paintbrush pattern on the capsular tissue in the chamber and a fluoroptic thermometry probe was placed 1 cm above the RFE treatment probe to record the fluid temperature. Both intermittent and the continuous treatment manners were tested under flow and no-flow conditions. For each probe/manner/flow combination, 6 bovine capsular tissue specimens were tested (n = 6). All data were recorded using a HyperTerminal software program (Hilgraeve Inc, Monroe, MI) into a personal computer.ResultsWhen using intermittent and continuous treatment manners with flow, all recorded chamber fluid temperatures for all tested RFE probes at each time interval were below 40 degrees C. Under no-flow conditions, with intermittent treatment, the ArthroCare probe caused joint fluid temperatures to exceed 50 degrees C after 70 seconds of RFE treatment. With the continuous treatment, the ArthroCare caused chamber fluid temperatures to exceed 65 degrees C after 2 minutes of treatment. The highest mean recorded chamber fluid temperature was caused by ArthroCare probe, which reached 80 degrees C at 3 minutes. For all probes, continuous treatment caused significantly higher chamber fluid temperatures than intermittent treatment.ConclusionsThe results of this study indicate that using flow during thermal capsulorrhaphy could lower joint fluid temperature to prevent heated joint fluid from killing chondrocytes of articular cartilage, and the intermittent treatment manner caused lower fluid temperature compared with continuous treatment within the RFE-treated shoulder joint.Clinical RelevanceArticular cartilage of the humeral head may suffer potential thermal injury from heating of joint fluid during RFE thermal capsulorrhaphy.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.