-
- B Milne, F W Cervenko, M B Henderson, and P J Westra.
- Can Anaesth Soc J. 1986 Mar 1;33(2):234-6.
AbstractDuring a coronary artery bypass operation arterial blood pressure measured with a Bentley Trantec model 800 transducer increased erroneously while continuous electrocautery was being used. This phenomenon has recurred infrequently, with fictitious hypotension being observed in one patient. To reproduce the problem of pressure offset during electrosurgery a bench test demonstrated that with peak to peak voltage of 20 volts from the electrosurgical unit, three of seven Bentley transducers had offsets as much as +/- 50 mmHg. It is important for anaesthetists to determine if electrosurgery units are functioning before treating apparent pressure drifts.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.