• Journal of neurosurgery · Jan 1999

    Accuracy of true frameless stereotaxy: in vivo measurement and laboratory phantom studies. Technical note.

    • N L Dorward, O Alberti, J D Palmer, N D Kitchen, and D G Thomas.
    • University Department of Neurosurgery, Institute of Neurology, National Hospital for Neurology and Neurosurgery, London, United Kingdom.
    • J. Neurosurg. 1999 Jan 1;90(1):160-8.

    AbstractThe authors present the results of accuracy measurements, obtained in both laboratory phantom studies and an in vivo assessment, for a technique of frameless stereotaxy. An instrument holder was developed to facilitate stereotactic guidance and enable introduction of frameless methods to traditional frame-based procedures. The accuracy of frameless stereotaxy was assessed for images acquired using 0.5-tesla or 1.5-tesla magnetic resonance (MR) imaging or 2-mm axial, 3-mm axial, or 3-mm helical computerized tomography (CT) scanning. A clinical series is reported in which biopsy samples were obtained using a frameless stereotactic procedure, and the accuracy of these procedures was assessed using postoperative MR images and image fusion. The overall mean error of phantom frameless stereotaxy was found to be 1.3 mm (standard deviation [SD] 0.6 mm). The mean error for CT-directed frameless stereotaxy was 1.1 mm (SD 0.5 mm) and that for MR image-directed procedures was 1.4 mm (SD 0.7 mm). The CT-guided frameless stereotaxy was significantly more accurate than MR image-directed stereotaxy (p = 0.0001). In addition, 2-mm axial CT-guided stereotaxy was significantly more accurate than 3-mm axial CT-guided stereotaxy (p = 0.025). In the clinical series of 21 frameless stereotactically obtained biopsies, all specimens yielded the appropriate diagnosis and no complications ensued. Early postoperative MR images were obtained in 16 of these cases and displacement of the biopsy site from the intraoperative target was determined by fusion of pre- and postoperative image data sets. The mean in vivo linear error of frameless stereotactic biopsy sampling was 2.3 mm (SD 1.9 mm). The mean in vivo Euclidean error was 4.8 mm (SD 2 mm). The implications of these accuracy measurements and of error in stereotaxy are discussed.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…