Paediatric anaesthesia · Jan 2000
Comparative StudyImpact of infusion line compliance on syringe pump performance.
- M Weiss, T Neff, A Gerber, and J Fischer.
- Department of Anaestesia, University Children's Hospital of Zurich, Switzerland.
- Paediatr Anaesth. 2000 Jan 1;10(6):595-9.
AbstractTo determine the impact of infusion line compliance on the time to occlusion alarm (occlusion time) and the size of the subsequent occlusion release bolus, we studied three different infusion lines in combination with a 50-ml and 10-ml syringe. The mean occlusion time of the 50-ml syringe amounted to 20.3 +/- 0.28 min and increased to 26.2 +/- 0.19 min with the infusion with the infusion line yeielding the largest compliance (P<0.0001). The mean occlusion time of the 10-ml syringe was 2.5 +/- 0.75 min and showed a 2.7-fold variation with the three infusion lines (2.6+/-0.08, 5.6 +/- 0.28 and 6.9 +/- 0.35 min, respectively; P<0.0001). Occlusion times correlated strongly to be calculated estimated compliance of the syringe-infusion line assembly (Pearson's r=0.998, P<0.0001). Infusion line compliance has an important impact on time to occlusion, particularly when using small, low compliant infusion pump syringes.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.