Revista de neurologia · Sep 2013
Observational Study[High performance of an implantable Holter monitor in the detection of concealed paroxysmal atrial fibrillation in patients with cryptogenic stroke and a suspected embolic mechanism].
- Esther Rojo-Martinez, María Sandín-Fuentes, Ana I Calleja-Sanz, Elisa Cortijo-García, Pablo García-Bermejo, Marina Ruiz-Piñero, Jerónimo Rubio-Sanz, and Juan F Arenillas-Lara.
- Hospital Clinico Universitario de Valladolid, 47005 Valladolid, Espana. erojo80@yahoo.it
- Rev Neurol. 2013 Sep 16;57(6):251-7.
IntroductionImplantable loop recorders (ILR) may allow detection of occult paroxysmal atrial fibrillation (PAF) in patients with cryptogenic ischemic stroke. However, optimal selection algorithm and ideal duration of monitoring remain unclear. AIM. To determine the incidence and time-profile of PAF in patients with cryptogenic ischemic stroke studied with Reveal XT ILR, who were selected based on a high suspicion of cerebral embolism.Patients And MethodsSelection Criteriaabsence of stroke etiology after complete study including vascular imaging, transesophageal echocardiography and at least 24 hours of cardiac rhythm monitoring, and confirmation of acute embolic occlusion of intracranial artery by transcranial duplex or characteristics of acute ischemic lesion on neuroimaging suggesting embolic mechanism of ischemia. After implanting Reveal XT ILR, patients were trained to perform transmissions monthly or when symptoms occurred. We reviewed the information online each month and patients underwent clinical visits at 3rd and 6th month and then every six months.ResultsWe included 101 patients with cryptogenic ischemic stroke and at least one month of follow-up after ILR implant. Mean age was 67 years, 54 women (53.5%). Mean follow-up after implantation was 281 ± 212 days. Occult PAF was detected in 34 patients (33.7%). Frequency of false positives: 22.8%. Median time from implant to arrhythmia detection was 102 days (range: 26-240 days). 24 (70%) patients with PAF had several arrhythmic episodes detected with ILR. The majority of events (75%) were detected during the first six months of monitoring.ConclusionsIn our patients with probably embolic cryptogenic ischemic stroke, PAF was detected by Reveal XT ILR in 33.7%. One in four events occurred after the first six months of monitoring.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.