In vitro evaluation of hydraulic characteristics of prototype implantable intrathecal infusion pump.
- Kyoung Won Nam, Mun Hyun Sung, Hyun Ho Kim, Kwang Gi Kim, Dae Hyun Kim, Byeong Han Lee, and Yung Ho Jo.
- Biomedical Engineering Branch, National Cancer Center, Goyang, Gyeonggi-do, Korea.
- Int J Artif Organs. 2010 Dec 1;33(12):895-9.
AbstractThe use of the intrathecal infusion pump for therapeutic treatment and pain management is increasing. For example, one such application is the pain treatment of cancer patients suffering from severe chronic pain, where all other treatment methods have failed. This method is gaining popularity because of its high cure effect with low dosage. In this study, we developed a prototype implantable intrathecal infusion pump and evaluated its mechanical and hydraulic characteristics in vitro to determine how its performance varied under different environmental conditions. The data are reported as means (standard deviations). In the experiments, the prototype pump could control the micro-scale infusion amount, and its performance was affected by ambient temperature and pressure conditions. In a temperature change test, at a constant pressure of 1.0 atm, the minimal amounts of a bolus were 4.44 (1.07), 5.06 (1.17), and 5.54 (0.90) uL for the temperature of 27.5, 36.5, and 42°C, respectively. In a pressure change test, at a constant temperature of 36.5°C, the minimal amounts of a bolus were 5.06 (1.17), 5.94 (0.67), and 6.13 (0.39) uL for pressures of 1.0, 0.9 and 0.8 atm, respectively. These experimental results demonstrate the possibility of using the prototype pump as an implantable microvolumetric infusion device. However, this prototype pump will have to undergo further design enhancement before being clinically feasible for such an application.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.