- R J Taylor, G Smurthwaite, I Mehmood, G B Kitchen, and R D Baker.
- Department of Medical Physics, Salford Royal NHS Foundation Trust (SRFT), Salford, UK.
- Anaesthesia. 2015 Jan 1; 70 (1): 18-25.
AbstractWe describe the development and laboratory assessment of a refined prototype tactile feedback device for the safe and accurate application of cricoid pressure. We recruited 20 operating department practitioners and compared their performance of cricoid pressure on a training simulator using both the device and a manual unaided technique. The device significantly reduced the spread of the applied force: average (SE) root mean squared error decreased from 8.23 (0.48) N to 5.23 (0.32) N (p < 0.001). The average (SE) upwards bias in applied force also decreased, from 2.30 (0.74) N to 0.88 (0.48) N (p < 0.01). Most importantly, the percentage of force applications that deviated from target by more than 10 N decreased from 18% to 7% (p < 0.01). The device requires no prior training, is cheap to manufacture, is single-use and requires no power to operate, whilst ensuring that the correct force is always consistently applied. © 2014 The Association of Anaesthetists of Great Britain and Ireland.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.