• Reg Anesth Pain Med · Nov 2014

    Evaluation of Epidural and Peripheral Nerve Catheter Heating During Magnetic Resonance Imaging.

    • Sean Owens, M Arcan Erturk, Jean-Pierre P Ouanes, Jamie D Murphy, Christopher L Wu, and Paul A Bottomley.
    • From the *Department of Anesthesiology and Critical Care Medicine and †Division of MR Research, Russell H. Morgan Department of Radiology and Radiologic Science, The Johns Hopkins University and School of Medicine, Baltimore, Maryland.
    • Reg Anesth Pain Med. 2014 Nov 1; 39 (6): 534-9.

    BackgroundMany epidural and peripheral nerve catheters contain conducting wire that could heat during magnetic resonance imaging (MRI), requiring removal for scanning.MethodsWe tested 2 each of 6 brands of regional analgesia catheters (from Arrow International [Reading, Pennsylvania], B. Braun Medical Inc [Bethlehem, Pennsylvania], and Smiths Medical/Portex [Keene, New Hampshire]) for exposure to clinical 1.5- and 3-T MRI. Catheters testing as nonmagnetic were placed in an epidural configuration in a standard human torso-sized phantom, and an MRI pulse sequence applied at the maximum scanner-allowed radiofrequency specific absorption rate (SAR) for 15 minutes. Temperature and SAR exposure were sampled during MRI using multiple fiberoptic temperature sensors.ResultsTwo catheters (the Arrow StimuCath Peripheral Nerve and B. Braun Medical Perifix FX Epidural) were found to be magnetic and not tested further. At 3 T, exposure of the remaining 3 epidural and 1 peripheral nerve catheter to the scanner's maximum RF exposure elicited anomalous heating of 4°C to 7°C in 2 Arrow Epidural (MultiPort and Flex-Tip Plus) catheters at the entry points. Temperature increases for the other catheters at 3 T, and all catheters at 1.5 T were 1.4°C or less. When normalized to the body-average US Food and Drug Administration guideline SAR of 4 W/kg, maximum projected temperature increases were 0.1°C to 2.5°C at 1.5 T and 0.7°C to 2.7°C at 3 T, except for the Arrow MultiPort Flex-Tip Plus catheter at 3 T whose increase was 14°C.ConclusionsMost but not all catheters can be left in place during 1.5-T MRI scans. Heating of less than 3°C during MRI for most catheters is not expected to be injurious. While heating was lower at 1.5 T versus 3 T, performance differences between products underscore the need for safety testing before performing MRI.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…