- E K Paulson, D H Sheafor, D S Enterline, H P McAdams, and T T Yoshizumi.
- Department of Radiology, Duke University Medical Center, Erwin Rd, Box 3808, Durham, NC 27710, USA. pauls003@mc.duke.edu
- Radiology. 2001 Jul 1;220(1):161-7.
PurposeTo determine the radiation dose to radiologists who perform computed tomographic (CT) fluoroscopic interventional procedures by using a quick-check method and a low-milliampere technique.Materials And MethodsTwo hundred twenty CT fluoroscopy--guided interventional procedures were performed in 189 patients. Procedures included 57 spinal injections, 17 spinal biopsies, 24 chest biopsies, 20 abdominal aspirations, 44 abdominal biopsies, and 58 abdominal drainages. Procedure details were prospectively recorded and included site, depth, target diameter, milliampere value, kilovolt peak, fluoroscopic time, and CT technique (continuous CT fluoroscopy, quick-check method, or a combination of these techniques). An individual collar and finger radiation detector were worn by each radiologist during each procedure to determine the dose per procedure.ResultsThe quick-check technique was performed in 191 (87%) of 220 procedures. Four procedures were performed with continuous CT fluoroscopy, and a combination technique was used for 25 (11%) procedures. The overall mean CT fluoroscopic time was 17.9 seconds (range, 1.2--101.5 seconds). The mean milliampere value was 13.2 mA (range, 10--50 mA). The overall mean radiologist radiation dose per procedure was 2.5 mrem (0.025 mSv) (whole body). Individual procedure doses ranged from 0.66 to 4.75 mrem (0.007--0.048 mSv). The finger radiation dose was negligible.ConclusionBy using a low-milliampere technique and the quick-check method, CT fluoroscopic time and radiation exposure can be minimized.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.