Comparative StudyA comparison of two different bronchial cuff designs and four different bronchial cuff inflation methods.
- D C Guyton, T R Besselievre, M Devidas, L G DeLima, and J H Eichhorn.
- Department of Anesthesiology, University of Mississippi Medical Center, Jackson 39216-4505, USA.
- J. Cardiothorac. Vasc. Anesth. 1997 Aug 1;11(5):599-603.
ObjectiveTo compare directly measured pressures at the cuff/trachea interface that are associated with two different bronchial cuff designs and four different methods of bronchial cuff inflation suggested for use with one-lung ventilation.DesignIn vitro study.SettingExperimental laboratory in a university-affiliated hospital.InterventionsThe bronchial cuffs of two different endotracheal tubes were inflated using one of four different methods of determining the cuff volume and pressure necessary to "just seal" the bronchus and obtain lung separation; positive-pressure test, negative-pressure test, CO2 analysis, and a new test using an anesthesia ventilator. When each method predicted the "just-seal" state, the pressure at the cuff/bronchus interface as well as cuff inflation pressure and volume were recorded.Measurements And Main ResultsAlthough the new test was incompatible with the bronchial blocker, the other three tests all accurately predicted lung separation with the Univent, with no significant differences in pressures exerted on the trachea. However, when used with the double-lumen tube, the new test produced significantly lower measured parameters than the other inflation methods. CO2 analysis did not reliably predict bronchial seal with the double-lumen tube.ConclusionsTo achieve lung separation with the lowest pressure on the trachea, the new test is the most appropriate method for use with the double-lumen tube; however, the negative-pressure test appeared to be the easiest and fastest method for use with the bronchial blocker. Although the bronchial blocker was associated with lower pressures transmitted from the cuff to the trachea, the in vitro model cannot predict which bronchial cuff design would be superior in vivo.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.