    Indicator amount, temperature, and intrinsic cardiac output affect thermodilution cardiac output accuracy and reproducibility.

    • L E Renner, M J Morton, and G Y Sakuma.
    • School of Nursing, Department of Medicine, Oregon Health Sciences University, Portland 97201-3098.
    • Crit. Care Med. 1993 Apr 1;21(4):586-97.

    ObjectiveTo determine the accuracy and reproducibility of four thermodilution indicators (5-mL room temperature, 10-mL room temperature, 5-mL iced, and 10-mL iced injectates) at clinically relevant flow rates.DesignQuasi-experimental study.SettingAnimal research laboratory of a health sciences university.SubjectsSix virgin western-breed ewes.InterventionsData were collected from six ewes that had ascending aorta electromagnetic flow probes and inferior vena cava occluders. Cardiac output was manipulated by inferior vena cava occlusion and isoproterenol infusion. Four thermodilution indicators were tested at high and low levels of cardiac output and compared with the electromagnetic flowmeter measurements of cardiac output.Measurements And Main ResultsThe indicator amounts were determined from both injectate volume and temperature difference between the injectate and blood. Using 5-mL room temperature injectate as a reference, 10-mL room contained 2 x, 5-mL iced 2.1 x, and 10-mL iced 4.1 x the indicator amount of 5-mL room temperature injectate. Approximately 210 simultaneous thermodilution and electromagnetic flow measurements were made for each injectate over a flow range of 1.5 to 15.7 L/min. For the entire cardiac output range, systematic error was not present. However, the r2 value (.92) for the 10-mL iced injectate group was greater (p < .05) than that value (.79) for the 5-mL iced injectate group, while r2 values were .79 for the 10-mL room temperature group and .49 for the 5-mL room temperature group. At flow rates of < 4.7 L/min, r2 was not different among injectates, but reduced indicator amounts progressively overestimated output (p < .05), reaching 21% for the 5-mL room temperature group. At flow rates of > 7.7 L/min, the r2 value (.81) for the 10-mL iced group was greater (p < .05) than that value (.45) for the 5-mL iced group, while r2 values were .24 for the 10-mL room temperature group and .08 for the 5-mL room temperature group. Systematic error was not present.ConclusionsAt low cardiac output levels, reduced indicator impairs accuracy but not reproducibility, a phenomenon that is perhaps related to indicator loss. At high cardiac output rates, reduced indicator impairs reproducibility. This phenomenon is probably related to low signal-to-noise ratio. Thermodilution indicator amounts should be tailored to the output range.

      Pubmed     Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

Want more great medical articles?

Keep up to date with a free trial of metajournal, personalized for your practice.
1,694,794 articles already indexed!

We guarantee your privacy. Your email address will not be shared.