• Am. J. Crit. Care · Nov 2004

    Equivalence of the bioimpedance and thermodilution methods in measuring cardiac output in hospitalized patients with advanced, decompensated chronic heart failure.

    • Nancy M Albert, Melanie D Hail, Jianbo Li, and James B Young.
    • Division of Nursing, Cleveland Clinic Foundation, Cleveland, Ohio, USA.
    • Am. J. Crit. Care. 2004 Nov 1; 13 (6): 469-79.

    BackgroundAn accurate and reliable noninvasive method for determining cardiac output/cardiac index would be valuable for patients with acutely decompensated advanced systolic heart failure.ObjectivesTo determine whether a correlation exists for cardiac output and index determined by using bioimpedance and thermodilution in patients with acutely decompensated complex heart failure and if differences between results with the 2 methods could be explained by the patients' advanced condition.MethodsCardiac output and index were determined by using bioimpedance and thermodilution in 33 patients. Echocardiographic and electrocardiographic data were assessed to determine if differences between results with the 2 methods could be explained by the patients' advanced condition. Concordance correlation coefficients and Bland-Altman agreement between methods were calculated.ResultsFour patients were excluded from analysis because reliable measurements could not be obtained; the remaining 29 patients constituted the study population. Mean cardiac outputs determined by thermodilution and bioimpedance were 5.48 and 5.40 L/min, respectively (rhoc = 0.89, P < .001), and mean cardiac indexes were 2.67 and 2.65 (rhoc = 0.82, P < .001). Mean bias (limits of agreement) between data pairs was 0.08 (-0.18 to 0.35) L/min (P = .52) for cardiac output and 0.03 (-0.097 to 0.16; P = .61) for cardiac index. Six data pairs (21%) had an absolute percent difference greater than 15%. Of these, 50% had a higher thermodilution value.ConclusionDeterminations of cardiac output and index by both methods were significantly correlated. Mean bias between the 2 methods was small, suggesting clinical utility for bioimpedance in patients with complex decompensated heart failure.

      Pubmed     Free full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…