• Acta neurochirurgica · Jul 2009

    Contact position analysis of deep brain stimulation electrodes on post-operative CT images.

    • Simone Hemm, Jérôme Coste, Jean Gabrillargues, Lemlih Ouchchane, Laurent Sarry, François Caire, François Vassal, Christophe Nuti, Philippe Derost, Franck Durif, and Jean-Jacques Lemaire.
    • Inserm, ERI 14, Equipe de Recherche en Imagerie Médicale, Clermont-Ferrand 63001, France.
    • Acta Neurochir (Wien). 2009 Jul 1; 151 (7): 823-9; discussion 829.

    PurposeGroups performing deep brain stimulation advocate post-operative imaging [magnetic resonance imaging (MRI) or computer tomography (CT)] to analyse the position of each electrode contact. The artefact of the Activa 3389 electrode had been described for MRI but not for CT. We undertook an electrode artefact analysis for CT imaging to obtain information on the artefact dimensions and related electrode contact positions.MethodsThe electrode was fixed on a phantom in a set position and six acquisitions were run (in-vitro study). The artefacts were compared with the real electrode position. Ten post-operative acquisitions were analysed (in-vivo analysis). We measured: H (height of the lateral black artefact), D (distance between the beginning of the white and the lateral black artefacts) and W (maximal artefact width), representing respectively the lengths of the four contacts and the electrode tip and width of the contact zone. A Student t-test compared the results: in vivo vs in vitro and coronal vs sagittal reconstructions along the electrode.ResultsThe limits of the lateral black artefact around the electrode contacts corresponded to the final electrode position. There was no significant difference for D (in vivo, 1.1 +/- 0.1 mm; in vitro, 1.2 +/- 0.2 mm; p = 0.213), while W and H differed slightly (in vivo, W = 3.3 +/- 0.2 mm, H = 7.7 +/- 0.2 mm; in vitro, W = 3.1 +/- 0.1 mm, H = 7.5 +/- 0.2 mm). Results obtained with sagittal and coronal reconstructions were similar (p > 0.6).ConclusionsPrecise three-dimensional (3D) localisation of the four-contact zone of the electrode can be obtained by CT identification of the limits of the lateral black artefact. The relative position of the four contacts is deduced from the size of the contacts and the inter-contact distance. Sagittal and coronal reconstructions along the electrode direction should be considered for the identification of the four electrode contacts. CT offers a useful alternative to post-operative MRI.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…